It was reported that during a reconstruction of anterior cruciate ligament, at the end of the procedure, just at the moment of fixing the graft in the femoral tunnel, an interference screw (biorci 7 * 25, 8 mm head) was used.This was being screwed properly, but at the end of the process, the screwdriver was rolled inside the head's screw, and this did not advance more from that site.Surgeon tried to return it but this was not possible, screwdriver was removed, and it was noticed that the screw was broken and that it was not possible to move it forward or back.Due to this, the remaining screw head had to be broken to remove the fragment.No patient injury was reported.
|
|
Examination is not possible, as the device will not be returned, however a photograph was provided.The photo shows the head of a biorci screw that has broken off.Examination of the photograph is inconclusive.The investigation could not draw any conclusions about the reported event without the return of the device.
|