Model Number 2000US
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient's shower bag had a damaged strap.The shower bag was replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the shower bag was returned for evaluation.Failure analysis of the shower bag revealed a damaged strap.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed, but not limited, to wear and/or due to the handling of the bag.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - SHOWER BAG
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7820429
Report Number3007042319-2018-04106
Device Sequence Number1809022
Product Code DSQ
UDI-Device Identifier00888707000031
UDI-Public00888707000031
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2019
Device Model Number2000US
Device Catalogue Number2000US
Device Lot Number1417392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2018
Supplement Date Received by Manufacturer12/05/2018
Initial Report FDA Received Date08/27/2018
Supplement Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
-
-