Model Number MMT-523NAB
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had keypad anomaly.The customer also stated that they had button error alarm.The customer¿s blood glucose level was 215 mg/dl.The customer stated that pump got displayable history crc check failed on startup.Troubleshooting was done for keypad anomaly.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Insulin pump had pump history crc failed.Alarm in alarm history file.Pump history crc failed.Alarms due to corrupted history files.Insulin pump passed displacement test, rewind test, basic occlusion test, prime/a33 test, excessive no delivery test and occlusion test.No motor error alarm noted.The motor was tested outside of the device and passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7821668
Report Number2032227-2018-19475
Device Sequence Number1510869
Product Code OYC
UDI-Device Identifier00643169202306
UDI-Public(01)00643169202306
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAB
Device Catalogue NumberMMT-523NAB
Device Lot NumberA8523NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2018
Initial Date Received by Manufacturer 08/01/2018
Supplement Date Received by Manufacturer12/07/2018
Initial Report FDA Received Date08/28/2018
Supplement Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age16 YR
Patient Weight144
-
-