Model Number AR40E
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: the exact date is unknown, but the year 2017 was provided.If implanted; give date: unknown if the lens was implanted.If explanted; give date: unknown if the lens was implanted and therefore unknown if explanted.(b)(6).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that during implantation of the intraocular lens (iol) the patient's posterior capsule was broken by the lens lead haptic.It was also noted that the lens was defective.No additional information was provided.The same issue was reported for 2 lenses, therefore 2 mdrs will be filed.This report is the second of 2.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 10/05/2018.Device returned to manufacturer? yes.Device evaluation: the ar40e lens was returned for investigation.Visual inspection using magnification was performed where loose fibers/particles were observed on lens and related to handling of the unit out of a sterile environment.Viscoelastic residue was observed on lens.Both lens haptics were observed slightly distorted.The condition in which the sample was returned is consistent with a lens that was handled and prepared for a surgical process.No damage that could cause a tear in the lens was identified.The complaint issue report of capsule tear could not be verified, but slightly distorted, damaged haptics were verified and related to handling.Manufacturing record review: the manufacturing process record was evaluated and no deviation or discrepancy was found.The product was manufactured and released according to specifications.A search revealed that no additional investigations were requested for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-(b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7822017
Report Number2648035-2018-01160
Device Sequence Number19118372
Product Code HQL
UDI-Device Identifier05050474502192
UDI-Public(01)05050474502192(17)220626
Combination Product (Y/N)Y
Initial Reporter CountryES
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/30/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2022
Device Model NumberAR40E
Device Catalogue NumberAR40E00245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/31/2018
Supplement Date Received by Manufacturer10/19/2018
10/25/2020
Initial Report FDA Received Date08/28/2018
Supplement Report FDA Received Date11/07/2018
10/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA01/01/2005
-
-