• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4471
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
The homechoice device was returned and evaluated by the product analysis lab.An internal/external inspection was performed and the device passed, along with all of the electrical testing.Returned instrument testing evaluation (rite) functional testing failed for fluid transferred outside of the specified limits, and the volumetric accuracy test that followed also failed.The membrane gasket was removed and revealed the original piston foam is deteriorated.The results of the evaluation revealed the cause of the failure to be the deteriorated piston foam.The piston foam will be replaced to resolve the reported issue.The device was sent for servicing.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, a baxter technician determined the device failed homechoice return instrument test/evaluation (rite) functional testing for fluid volume accuracy.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
largo FL
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7822251
MDR Text Key118469495
Report Number1416980-2018-05441
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4471
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Initial Date Manufacturer Received 08/08/2018
Initial Date FDA Received08/28/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-