TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Codes Unable to Obtain Readings (1516); Ambient Temperature Problem (2878)
Health Effect - Clinical Codes Dizziness (2194); Shaking/Tremors (2515)
Date of Event 08/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-33-ambient temperature out of operating range test strips udi# (b)(4.Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.
 
Event or Problem Description
Consumer reported complaint for e-1 error code accompanied by symptom.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of shakiness and dizziness however customer does not wish to seek medical attention.Medical attention is not reported as a result of the actual symptoms.The product storage location is undisclosed.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test results of e-1 using true metrix meter.The test strip lot manufacturer's expiration date is 07/29/2019 and open vial date is date.The meter memory was not reviewed for previous test result history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7822444
Report Number1000113657-2018-00895
Device Sequence Number19155889
Product Code NBW
UDI-Device Identifier00052569138557
UDI-Public(01)00052569138557
Combination Product (Y/N)N
Initial Reporter StateAR
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2766
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2018
Initial Report FDA Received Date08/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-