| Model Number |
VSR100812P |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter during a laparoscopic hysterectomy, when the surgeon moved in and out of the mirror viewing needle, it caught on the valve of the device and the device got out of the abdominal wall.The procedure was completed with another device.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted: only the trocar, circular and envelope seal were received.No abnormalities were noted.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|