Blank fields on this form indicate the information is unchanged, unknown or unavailable.Investigation - evaluation a review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.The proper procedures are in place to identify and prevent this failure mode prior to device distribution.The instructions for use (ifu) instruct the user to inspect the product prior to use to ensure that no damage has occurred.A review of the device history record was performed, revealing one nonconformance for the curve of the catheter not aligning with the template.Nonconformances were also reported for the catheter tubing subassembly lots for incorrect work orders, embedded foreign matter and insufficient hydrophilic coverage.All affected devices were scrapped out or reworked prior to device release.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no returned product and the results of our investigation, a definitive cause could not be established.Appropriate measures have been taken to address this failure mode.Per the quality engineering risk assessment no further action is required.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|