| Model Number |
H7493926024350 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).Device evaluated by mfr: synergy ii us mr 3.50 x 24mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent identified stent damage.The mid-section stent row was damaged with stent struts being damaged and lifted from the stent profile.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found multiple kinks along the full length of the catheter.A visual and tactile examination of shaft polymer extrusion revealed no issues.A visual and microscopic examination found damage to the tip.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 03-aug-2018.It was reported that crossing difficulties were encountered.The target lesion was located in a coronary artery.A 3.50 x 24 synergy ii drug-eluting stent was advanced but failed to cross the lesion.No patient complications were reported.However, returned device analysis revealed stent damage.
|
| |
|
Search Alerts/Recalls
|