| Catalog Number |
151650306 |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Codes |
Pain (1994); Not Applicable (3189); No Code Available (3191)
|
| Date of Event |
08/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised to address pain and instability due to poly size.Doi: (b)(6) 2018; dor: (b)(6) 2018; left knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|