| Model Number |
2008K@HOME |
| Medical Device Problem Code |
Inadequate Ultra Filtration (1656)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
| |
|
Event or Problem Description
|
|
Hemodialysis clinic contact reported that the customer reports a low ultra filtration removal during treatment.Follow up with the clinic indicates that the patient did not require medical intervention.The uf goal was 3.0 but the machine pulled 1.5.However, the patient lost weight during the last checkup.The biomed technician was scheduled to test the machine with simulation treatments.There was no reported malfunction or alarms to the machine.Additional information solicited.
|
| |
|