| Model Number |
8393249 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Perforation (2001); Bowel Perforation (2668)
|
| Date of Event |
07/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The patient remained in hospital for longer than was expected.Rwmic will attempt to obtain additional information from (b)(4) on patient demographics and other missing information.When evaluation results become available, a follow up report will be submitted as appropriate.Rwmic considers this report open at this time.
|
| |
|
Event or Problem Description
|
|
Richard wolf medical instruments corporation (hereinafter referred to as rwmic) was made aware of an event from richard wolf (b)(4) (hereinafter referred to as (b)(4)) on 8/17/2018.(b)(4) was notified on 7/31/2018 by distributor in (b)(6) of an event that occurred at the hospital on (b)(6) 2018 during a small bowel occlusion surgery.During procedure 7 perforations were noted and determined to be likely caused by manipulation with the forceps.The bowel was uncharacteristic during the procedure and the manipulation was not excessive.The doctor had to set several clamps which had to be removed later.The patient required to stay in hospital longer than expected.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to the manufacturer for evaluation.The examination of the 3 sent in internal parts showed that they product specifications and can be used thus without hesitation.The inner parts have no edges, chamfers or sharp contours.This indicates an error in the use of the device.Type (b)(4) is the inner part of an atraumatically shaped grasping forceps for gripping and manipulating small and large intestine tissue.The inner part of the forceps is designed in such a way that even very sensitive intestinal tissue can be manipulated safely and atraumatically at any time.The probability of the occurrence with regard to the occurrence is to be estimated as low, since so far no such occurrences have been reported to richard wolf gmbh in the last 4 years.When manipulating intestinal tissue, care should be taken not to exert too much pressure on the tissue through the grasping forceps in order to avoid possible injuries.The probability that a tissue injury caused by the product is caused by proper use is unlikely but possible.Richard wolf gmbh considers this case as closed.
|
| |
|
Search Alerts/Recalls
|