• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION KNEE; FEMUR, NONPOROUS, SIZE10, RT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION KNEE; FEMUR, NONPOROUS, SIZE10, RT Back to Search Results
Model Number 202-01-110
Health Effect - Clinical Code Fall (1848)
Date of Event 08/14/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient falling and beginning to loosen.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to loosening.The previous surgery and the revision detailed in this investigation occurred 2.7 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There was a non-conforming material report (ncmr) # 20291 associated with the part # 202-01-110, femur, non porous, size 10,right, which documents a nonconformance that out of 9 quantity, 5 items were rejected due to out of tolerance of a perpendicularity tolerance.It was reworked with suitable operation and were accepted.All other items in the lot met the design, fit and function requirements.The device was within its respective expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.Agent has clearly mentioned that the patient fell and began to loosen, it seems that the event may have been occurred due to trauma, patient noncompliance with medical instructions or patient activities.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.
 
Additional Manufacturer Narrative
Additional narrative: concomitant medical products: additional concomitant items needed to be added to the mdr complaint.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
FEMUR, NONPOROUS, SIZE10, RT
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7824460
Report Number1644408-2018-00812
Device Sequence Number1143727
Product Code JWH
UDI-Device Identifier00888912109437
UDI-Public(01)00888912109437
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K923277
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2019
Device Model Number202-01-110
Device Catalogue Number202-01-110
Device Lot Number743A1035
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/14/2018
Supplement Date Received by Manufacturer09/07/2018
09/07/2018
Initial Report FDA Received Date08/28/2018
Supplement Report FDA Received Date09/07/2018
02/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
385-11-504, LOT 59601306; 402283, LOT 293720; 402283, LOT 405990; 385-11-504, LOT 59601306
Outcome Attributed to Adverse Event Required Intervention;
-
-