Model Number 313.97
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device used in a veterinary case - no patient information will be reported.Device is an instrument and is not implanted/explanted.Dhr review was completed.Please note, lot number 1849625 indicates part 313.940 which is the complete part including 313.970.Part: 313.940 , lot: 1849625.Manufacturing site: (b)(4).Release to warehouse date: 20.Mar.2008.The device history record shows this lot of 24 pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Service & repair evaluated the returned device.Service & repair evaluation the customer reported the holding sleeve was not working properly.The repair technician reported the spring was missing.Missing parts is the reason for repair.The cause of the issue is unknown.The following parts were replaced: spring.The item was repaired per the inspection sheet, passed synthes final inspection on (b)(6) 2018 and will be returned to the customer upon completion of the service and repair process.Attached service record router completed through operation 30.Finalized service record will be archived in document management system.The evaluation was confirmed.The device was deemed serviceable and returned to the customer.No design or manufacturing issues were identified; therefore, no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This report is for a veterinary case.There was no human patient involvement.It was reported that on an unknown date the holding sleeve was not functioning properly.There was no patient involvement.This report is for one (1) holding sleeve.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLDING SLEEVE
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7824461
Report Number2939274-2018-53524
Device Sequence Number16423164
Product Code HXX
UDI-Device Identifier10887587037196
UDI-Public(01)10887587037196
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number313.97
Device Catalogue Number313.97
Device Lot Number1849625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/28/2018
Initial Report FDA Received Date08/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-