On 8/3/2018 sunrise medical received additional information from account manager, (b)(4) regarding the incident.(b)(4) stated, "i was able to speak with the local (b)(6) branch and their service tech.It appears the end user broke one or multiple ribs, along with some facial contusions.As far as the chair; on (b)(6) 2018 (b)(6) ordered replacement caster parts.They shipped the parts to the end user due to remote location to have user install.The end user was unable to remove the original parts, so scheduled an appointment at (b)(6) to have the tech install, which he installed in (b)(6) 2018.Apparently there were hammer indentations all over the top of the swing arms the (b)(6) tech noticed prior to the install of the replacement forks, etc.The (b)(6) tech believed the end user (and his buddies) tried to beat the original forks out with a hammer, but was unsuccessful, as they didn't realize there is a retaining ring below the bottom housing bearing.The branch stated they do not have any of the front fork assemblies, and nothing appears to be left in the front swing arm.He did say there was an oily substance all over the inside of the swing arms, as if it was sprayed with (b)(4) or another lubricant.Appears this is from abuse and/or neglect and not a manufacture defect." it was confirmed to sunrise medical with photos provided by the dealer that the front casters were completely missing from the front swing arm assemblies.The arm assemblies clearly show the markings where the caster assemblies were hammered out by the end user.Since the dealer is unable to provide the missing caster assemblies for further evaluation, sunrise medical considers this product failure to be due to end user abuse.There will be no further investigation performed by sunrise medical.
|