| Model Number |
10927766 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer reported a bulge in the silicone segment of the tubing shortly after starting the infusion and after the pump alarmed.The tubing was connected to a patient.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report of a bulge in the silicone segment was confirmed.Visual inspection of the set showed that the silicone segment tubing had a ballooned bulge near the upper fitment.Functional testing showed no anomalies.The root cause of the balloon/bulge in the silicone segment was identified as an iv push medication or flush being executed below the pump without first clamping the tubing above the injection port.
|
| |
|
Event or Problem Description
|
|
The customer reported a bulge in the silicone segment of the tubing shortly after starting the infusion and after the pump alarmed.The tubing was connected to a patient.There was no patient harm.
|
| |
|