Model Number 10927766
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
 
Event or Problem Description
The customer reported a bulge in the silicone segment of the tubing shortly after starting the infusion and after the pump alarmed.The tubing was connected to a patient.There was no patient harm.
 
Additional Manufacturer Narrative
The customer's report of a bulge in the silicone segment was confirmed.Visual inspection of the set showed that the silicone segment tubing had a ballooned bulge near the upper fitment.Functional testing showed no anomalies.The root cause of the balloon/bulge in the silicone segment was identified as an iv push medication or flush being executed below the pump without first clamping the tubing above the injection port.
 
Event or Problem Description
The customer reported a bulge in the silicone segment of the tubing shortly after starting the infusion and after the pump alarmed.The tubing was connected to a patient.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS® PUMP MODULE ADMINISTRATION SET
Common Device Name
SET,ADMINISTRATION,INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key7824737
Report Number9616066-2018-01498
Device Sequence Number16537444
Product Code FPA
UDI-Device Identifier07613203019101
UDI-Public7613203019101
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K944320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/26/2021
Device Model Number10927766
Device Catalogue Number10927766
Device Lot Number18026852
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/28/2018
Supplement Report FDA Received Date01/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8100,8015, TD
Patient Weight75
-
-