MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL 13.2 -
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm micl13.2, -8.00 diopter, implantable collamer lens into the patients right eye (od) on (b)(6) 2018.On (b)(6) 2018 the lens was exchanged for a shorter length lens due to excessive vault.This exchange resolved the problem.
 
Additional Manufacturer Narrative
The reporter indicated that the surgeon implanted a 13.2mm micl13.2 implantable collamer lens of -8.00 diopter into the patient's right eye (od) on (b)(6) 2018.Secondary yag was performed for enlargement of pi.The lens was explanted on (b)(6) 2018 due to excessive vault and a shorter lens was later implanted on the same day in a different surgery.The problem was resolved.(b)(4).
 
Additional Manufacturer Narrative
Deice evaluation: lens was returned in liquid, in lens vial.Visual inspection found no visible damage to the lens.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7824747
Report Number2023826-2018-01291
Device Sequence Number16511219
Product Code MTA
UDI-Device Identifier00841542103251
UDI-Public00841542103251
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Model NumberMICL 13.2 -
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Initial Date Received by Manufacturer 07/31/2018
Supplement Date Received by Manufacturer09/20/2018
11/06/2018
Initial Report FDA Received Date08/28/2018
Supplement Report FDA Received Date10/17/2018
11/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
-
-