• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Catalog Number 97400-001371
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer¿s device and was unable to duplicate the reported issue.However, the customer showed a physio service representative a video of the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
 
Event or Problem Description
The customer contacted physio-control to report that their device displayed the ¿service needed¿ across the display during a patient event.This is indicative of a potential device lock-up.The customer advised that during a transport of a patient, after hitting a bump in the road, their device display went blank and then ¿service needed¿ appeared on the screen.When this occurred all leds became illuminated and remained lit.After approximately 5 minutes they powered the device off and then back on again, thereafter no further issues were observed.The customer advised that there were no adverse effects to the patient as a result of the reported issue.No further details were provided.Physio-control has made several attempts to contact the customer for further information on the event and patient, but has been unsuccessful.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7824772
Report Number0003015876-2018-01406
Device Sequence Number13846459
Product Code MKJ
UDI-Device Identifier00883873950982
UDI-Public00883873950982
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K102972
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number97400-001371
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2018
Initial Report FDA Received Date08/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-