| Model Number |
MMT-523LNAH |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/02/2013
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's father reported via phone call the insulin pump alarmed button error and cosmetic damage.The customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump reservoir chips away every time they try and lock it in place.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.The customer was advised to discontinue use of pump and revert to back-up plan.The customer's insulin pump is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm during testing noted.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime and excessive no delivery test due to buttons not responding.Insulin pump received with broken battery tube treads and broken reservoir tube lip.
|
| |
|