• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-RAD MEDICAL DIAGNOSTICS GMBH ANTI-HUMAN GLOBULIN (RABBIT/MURINE MONOCLONAL); ANTI-HUMAN-GLOBULIN, ANTI-IGG, -C3D; 10ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-RAD MEDICAL DIAGNOSTICS GMBH ANTI-HUMAN GLOBULIN (RABBIT/MURINE MONOCLONAL); ANTI-HUMAN-GLOBULIN, ANTI-IGG, -C3D; 10ML Back to Search Results
Catalog Number 804115100
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This is our combined initial and final report on this incident.
 
Event or Problem Description
The customer reported a false negative direct antiglobulin test (dat) of a cap survey sample when testing in the manual tube method with anti-human globulin anti-igg, -c3d; polyspecific.The customer returned the supposedly defective product for investigational testing and also the cap survey sample that had caused a false negative test result.Our quality control laboratory tested the cap survey sample with the complaint product sample returned by the customer and also with their retention sample in the direct antiglobulin test (dat) and yielded a correctly positive result.Additionally the complaint and the retention sample were tested for positive and negative specificity and potency.All acceptance criteria were met for both ahg reagents.We did not observe any false negative results.Testing by our quality control laboratory confirmed that the allegedly defective lot of anti-human globulin anti-igg, -c3d; polyspecific functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTI-HUMAN GLOBULIN (RABBIT/MURINE MONOCLONAL)
Common Device Name
ANTI-HUMAN-GLOBULIN, ANTI-IGG, -C3D; 10ML
Manufacturer (Section D)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM  63303
MDR Report Key7825709
Report Number9610824-2018-00054
Device Sequence Number11086103
Product Code KSZ
UDI-Device Identifier07611969950652
UDI-Public(01)07611969950652(17)181010(10)8630050-02
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2018
Device Catalogue Number804115100
Device Lot Number8630050-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-