| Catalog Number |
3L92506 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Concomitant med products: pinn mar +4 neut, 36idx56od; articuleze m head, 36mm, +8.5; corail amt collar, size 16; pinnacle sector ii cup, 56mm.
|
| |
|
Event or Problem Description
|
|
The patient was revised to address loosening of femoral stem at bone to implant interface and a retroverted cup.The surgeon removed all components.Doi: (b)(6) 2017; dor: (b)(6) 2018; unknown hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary :no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate. .
|
| |
|
Search Alerts/Recalls
|