DEPUY SPINE INC X-MESH PISTOL INSERTER/EXPANDE; DRIVER, PROSTHESIS Back to Search Results
Catalog Number 288900000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Injury (2348); Blood Loss (2597)
Date of Event 07/31/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A follow up report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported that a spinal surgery(not a revision) took place on (b)(6) 2018 where the following events happened with set # xmpc-40: 1.Part # 2889-00-000 lot br607613 cage could not be implanted as the inserter broke.2.Part #1889-01-233 lot bdc13sg failed to lock 3.Part #1889-01-028 lot bdcg3tk broke.4.Part # 1889-21-028 lot bdjf3sc would not release from driver.There were alternate parts available to complete the surgery.As a result of the reported issues, the patient consequences were as follows: 1.A prolonged surgery of 90 minutes.2.The vena cava was torn and had to be repaired.3.Loss of blood due to multiple attempts to complete the surgery.Surgeon report may be available in the near future.
 
Additional Manufacturer Narrative
(b)(4).It was noted that the expansion arms sub assembly was loose and was able to shift position/angle without input.Visual examination revealed that the expansion arms sub assembly was missing a screw which holds the arms of the sub assembly together.The screw was not returned to customer quality unit.There is no evidence of any breakage on the device.The expansion knob mechanism, release lever and pistol grip were functioning as intended.The gold sleeve which allowed the handles rotate and lock in at 90 degree intervals was functioning as intended.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for missing a screw which holds the arms of the sub assembly together cannot be positively determined with the information provided.However, noted damage suggests that the inserter/expande was likely subjected to unanticipated overloading forces while expanding the cage to the desired height, resulting in a screw which holds the arms of the sub assembly together falling apart.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-MESH PISTOL INSERTER/EXPANDE
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key7826070
Report Number1526439-2018-50816
Device Sequence Number16596345
Product Code HWR
UDI-Device Identifier10705034222592
UDI-Public(01)10705034222592
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number288900000
Device Lot NumberBR607613
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2018
Initial Date Received by Manufacturer 07/31/2018
Supplement Date Received by Manufacturer09/21/2018
Initial Report FDA Received Date08/29/2018
Supplement Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-