| Model Number |
FSC-3300-150-R |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The exact date of the event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event.The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy procedure performed on an unknown date.According to the complainant, during the procedure, the middle image was green.There were no patient complications reported as a result of this event.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The exact date of the event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event.A fuse scope was received for analysis.A functional evaluation was performed and there was no issue found in the device.The unit passed all testing.The image issue was not confirmed, the assigned investigation conclusion code for this complaint is designated as no problem detected.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy procedure performed on an unknown date.According to the complainant, during the procedure, the middle image was green.There were no patient complications reported as a result of this event.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|