| Model Number |
71962-05 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Blood Loss (2597)
|
| Date of Event |
08/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation is anticipated but has not yet begun.The hemoconcentrator sample was returned with tubing still attached.There was no damage or deformation observed during the visual inspection of exterior of the device.Next the device history record was reviewed and no issues were found related to this complaint.The 100% assessment for proper flow direction performed during the assembly of the convenience kit had been completed and no issues were identified.The supplier of the hemoconcentrator performs leak testing on 100% of the product during their manufacturing process.The supplier identified that the 100% fiber leak test is not always capable of detecting latent defects during longer periods of clinical use.The certificate of conformance (coc) was reviewed for this component and the component was deemed conforming by the supplier.The decontaminated sample has been forwarded to the supplier for further analysis.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that during a cardiopulmonary bypass (cpb) procedure, the hemoconcentrator fibers appeared to be broken.The hemoconcentrator was changed out.The surgical procedure was completed successfully.Minimal blood loss was reported.There was no delay nor adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up is submitted to the fda in accord with applicable regulations - and as indicated by terumo cardiovascular systems corporation in the initial report submitted august 29, 2018.Upon further investigation of this reported event, the following information is new and/or changed: identification of evaluation codes: method: 10 - testing of actual/suspected device; results: 115 - maintenance problem identified; conclusion 51 - cause traced to maintenance.The sample was forwarded to the supplier for further analysis.The sample was received by the supplier and the damaged fiber was confirmed during their evaluation.The supplier determined that all inventory of the affected lot had been sold and shipped, back on january 26, 2018.The root cause of this issue has been determined to be related to a gelled formation of undissolved pvp used in the manufacturing process of the fiber.The mixing tank is cleaned on an as-needed basis and there likely was a buildup due to the frequency of cleaning of the mixing tank.The supplier has introduced a routinely scheduled preventive maintenance for the cleaning of the mixing tank.Additionally, there was re-validation performed on the fiber leak test apparatus used during production of blood fiber products.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
|
| |
|
Search Alerts/Recalls
|