• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR(R) PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR(R) PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CRPLUS
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The customer received a replacement device.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
A customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and verified the reported issue.Physio-control determined that the cause of the reported issue was that hybrid layer capacitor (hlc) batteries, designators bt1, bt2 and bt3, of the analog pcb would no longer hold a charge.The hlc batteries would no longer hold a charge because they were damaged after exposure to water inside the device that resulted in corrosion of the internal components.
 
Event or Problem Description
A customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR(R) PLUS DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7827006
Report Number0003015876-2018-01408
Device Sequence Number16728605
Product Code MKJ
UDI-Device Identifier00885074440323
UDI-Public00885074440323
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCRPLUS
Device Catalogue Number99403-000156
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/14/2018
Supplement Date Received by Manufacturer10/01/2018
Initial Report FDA Received Date08/29/2018
Supplement Report FDA Received Date10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-