MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø10X300MM X 130°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø10X300MM X 130°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 34300300S
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available it will be provided on a supplemental report; hospital policy.
 
Event or Problem Description
Right knee revision - patient had a static knee spacer, a gamma nail and cement.Spacer was removed and her knee reconstructed using a gmrs distal femur construct with an mrh tibial baseplate.
 
Additional Manufacturer Narrative
The reported event that long nail kit r1.5, ti, right gamma3® ø10x300mm x 130° was alleged of issue (revision) could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use provides below mentioned adverse effects which can lead to revision surgery - delayed union or non-union of the fracture site.These devices can break when subjected to the increased loading associated with delayed unions and/or non-unions.Internal fixation devices are load sharing devices which are intended to hold fractured bone surfaces in apposition to facilitate healing.If healing is delayed or does not occur, the appliance may eventually break due to metal fatigue.Loads on the device produced by load bearing and the patient¿s activity level will dictate the longevity of the device.Conditions attributable to non-union, osteoporosis, osteomalacia, diabetes, inhibited revascularization and poor bone formation can cause loosening, bending, cracking, fracture of the device or premature loss of rigid fixation with the bone.Improper alignment can cause a mal-union of the bone and/or bending, cracking or even breakage of the device.¿ if device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
Right knee revision - patient had a static knee spacer, a gamma nail and cement.Spacer was removed and her knee reconstructed using a gmrs distal femur construct with an mrh tibial baseplate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø10X300MM X 130°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key7827321
Report Number0009610622-2018-01039
Device Sequence Number2434030
Product Code HSB
UDI-Device Identifier04546540375452
UDI-Public04546540375452
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date01/31/2022
Device Catalogue Number34300300S
Device Lot NumberK0D4A5D
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/02/2018
Supplement Date Received by Manufacturer10/11/2018
Initial Report FDA Received Date08/29/2018
Supplement Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-