• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 9/10 COCR HEAD 28MM +0; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 9/10 COCR HEAD 28MM +0; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 152190058
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Not Applicable (3189)
Date of Event 06/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The primary surgery was performed on 2005 via tha.It was reported that the revision surgery was performed on (b)(6) 2018 to revise the cup, the head, and the stem due to dissociation of the cup and the head.No further information was provided by the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9/10 COCR HEAD 28MM +0
Common Device Name
SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key7827469
Report Number1818910-2018-68047
Device Sequence Number1287559
Product Code JDI
UDI-Device Identifier10603295276630
UDI-Public10603295276630
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number152190058
Device Lot Number2519397
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/07/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
-
-