Model Number 45-480420
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was returned with the pusher for analysis; the coil was not attached to the pusher.The implant coil was slightly noted to be deformed, but still closely wound on the distal section.The implant coil was stretched on the proximal section, close to the tie-knot.The gel was noted to be absent on the proximal portion of the coil.There were no signs of damage to the pusher.The pusher was dissected to expose to analyze the attachment tether tip.The tip displayed minor signs of tensile break, and the rebound length was within specification.Based on the available information and evaluation of the returned device, the complaint can be confirmed.The root cause cannot be determined; however, the device exhibits evidence that it was subjected to tensile forces that exceeded its strength specifications.The instructions for use (ifu) identifies premature or difficult coil detachment as potential complications associated with use of the device.
 
Event or Problem Description
It was reported that during an embolization treatment, the coil detached partially in the microcatheter.The coil was removed in its entirety along with the microcatheter from the patient.There was no reported intervention or patient injury.The patient is reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR 18 PERIPHERAL V-TRAK SYSTEM
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7827516
Report Number2032493-2018-00170
Device Sequence Number2446892
Product Code HCG
UDI-Device Identifier00810170017926
UDI-Public(01)00810170017926(11)170308(17)220228(10)170308LC2
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K090168
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/28/2022
Device Model Number45-480420
Device Catalogue Number45-480420
Device Lot Number170308LC2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/30/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-