A us customer submitted their api escherichia coli survey strain ur-06 for ampicillin/sulbactam (ams) evaluation because susceptible results (mic=8) were obtained when the sample was tested on vitek® 2 ast-gn69 cards.The expected result for ams was resistant.Identification was confirmed, and testing included both the customer lot and a random lot of ast-gn69 cards.Agar dilution (ad), the reference method for ams (sam01n), was also performed.On the customer lot of ast-gn69 cards, a susceptible ams mic=8 was obtained, and, on the random lot, an intermediate mic=16 was obtained.Ad gave an intermediate mic=16.Both card results are within one (1) doubling dilution of reference method results, so card and reference method results are in agreement.The vitek 2 ast-gn69 cards performed as expected, and no further action is necessary.
|