Catalog Number 413400
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Event or Problem Description
A customer in the united states notified biomérieux of false susceptible results for an escherichia coli survey strain when testing with the vitek® 2 ast-gn69 test kit.The customer obtained a susceptible result (mic=8) for ampicillin/sulbactam.The expected result for this survey strain is resistant.As there is no patient associated with this quality control strain, there is no adverse event related to any patient's state of health.A biomérieux internal investigation will be initiated.
 
Additional Manufacturer Narrative
A us customer submitted their api escherichia coli survey strain ur-06 for ampicillin/sulbactam (ams) evaluation because susceptible results (mic=8) were obtained when the sample was tested on vitek® 2 ast-gn69 cards.The expected result for ams was resistant.Identification was confirmed, and testing included both the customer lot and a random lot of ast-gn69 cards.Agar dilution (ad), the reference method for ams (sam01n), was also performed.On the customer lot of ast-gn69 cards, a susceptible ams mic=8 was obtained, and, on the random lot, an intermediate mic=16 was obtained.Ad gave an intermediate mic=16.Both card results are within one (1) doubling dilution of reference method results, so card and reference method results are in agreement.The vitek 2 ast-gn69 cards performed as expected, and no further action is necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITEK® 2 AST-GN69 TEST KIT
Common Device Name
VITEK® 2 AST-GN69 TEST KIT
Manufacturer (Section D)
BIOMERIEUX INC.
595 anglum road
saint louis MO 63042
MDR Report Key7827713
Report Number1950204-2018-00312
Device Sequence Number1175903
Product Code LON
UDI-Device Identifier03573026398460
UDI-Public03573026398460
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
N50510: S 97
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2019
Device Catalogue Number413400
Device Lot Number5890677403
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2018
Supplement Date Received by Manufacturer10/14/2018
Initial Report FDA Received Date08/29/2018
Supplement Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-