• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ALTL ALANINE AMINOTRANSFERASE - PYRIDOXAL PHOPHATE ACTIVATED; NADH OXIDATION/NAD REDUCTION, ALT/SGPT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ALTL ALANINE AMINOTRANSFERASE - PYRIDOXAL PHOPHATE ACTIVATED; NADH OXIDATION/NAD REDUCTION, ALT/SGPT Back to Search Results
Catalog Number ASKU
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event occurred in: (b)(6).
 
Event or Problem Description
The customer originally complained about questionable crep2 creatinine plus ver.2 tested on both a cobas 6000 c (501) module and a cobas 8000 (unspecified module) compared to an abbott system.From the data provided, a reportable malfunction was provided for crep2, astlp aspartate aminotransferase acc.To ifcc with pyridoxal phosphate activation, altlp alanine aminotransferase acc.To ifcc with pyridoxal phosphate activation, and ggt-2 -glutamyltransferase ver.2 standardized against ifcc / szasz.This medwatch will cover altlp.Please refer to the medwatch with (b)(6) for information on astlp.Please refer to the medwatch with (b)(6) for information on ggt-2.Please refer to the medwatch with (b)(6) for information on crep2.The erroneous results were released outside of the laboratory.The patient was sent to the hospital due the initial high crep2 result, but there was no allegation that the patient was adversely affected.The cobas c501 serial (b)(4).The customer did not provide any specific information on the cobas 8000 system.A preliminary assessment of the case stated that based on previous electrophoresis testing of the patient, the patient is believed to have an interfering factor present.The investigation is currently ongoing.
 
Additional Manufacturer Narrative
The customer's cobas 8000 model used was a c701.The patient sample was received for further investigation.The results from the investigation have results indicative of a monoclonal gammopathy.Clarification was requested if this interference could have caused the discrepant results observed but clarification was not provided.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTL ALANINE AMINOTRANSFERASE - PYRIDOXAL PHOPHATE ACTIVATED
Common Device Name
NADH OXIDATION/NAD REDUCTION, ALT/SGPT
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7827930
Report Number1823260-2018-02916
Device Sequence Number16728593
Product Code CKA
UDI-Device Identifier04015630920532
UDI-Public4015630920532
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K951595
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/10/2018
Supplement Date Received by Manufacturer08/10/2018
Initial Report FDA Received Date08/29/2018
Supplement Report FDA Received Date10/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BACTRIM; INEXIUM; KARDEGIC; LEDERFOLINE; MILLEPERTUIS; PRETERAX; TAHOR; ZELITREX
Patient Age80 YR
-
-