MAUDE Adverse Event Report: MEDTRONIC XOMED INC. APS¿ (AUTOMATIC PERIODIC STIMULATION) ELECTRODE; ELECTRODE, NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. APS¿ (AUTOMATIC PERIODIC STIMULATION) ELECTRODE; ELECTRODE, NEEDLE Back to Search Results
Model Number 8228052
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis found the yellow c-clip contact was pulled back into the clip body.The contact of the returned sample was 0.07" away from the inside diameter, which was out of specification.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care professional (hcp) reported via a manufacturer representative that the probe had no signal during the thyroidectomy procedure.The procedure was delayed for about five minutes and another probe was used to detect better signals.This was the first time the probe was used.There was no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APS¿ (AUTOMATIC PERIODIC STIMULATION) ELECTRODE
Common Device Name
ELECTRODE, NEEDLE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7827992
Report Number1045254-2018-00400
Device Sequence Number11214794
Product Code GXZ
UDI-Device Identifier00643169309326
UDI-Public00643169309326
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083124
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2022
Device Model Number8228052
Device Catalogue Number8228052
Device Lot Number0215570615
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-