Model Number 186-0106 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/03/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the unit had low readings.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: one device was received and inspected.The sensor was tested and failed.An impedance check was performed and all electrodes passed the impedance check.The sensor was visually inspected- no physical anomalies were found, and witness marks were observed.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the unit had very low readings when the patient was sober.Customer reported that there was no patient injury.
|
|