| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from results obtained of 47 and 53mg/dl.The customer's expected fasting blood glucose test result range is 120 - 160mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer was not able to perform a back to back blood test during the call on (b)(6) /2018.The product is stored according to specification in the dining room.The test strip lot manufacturer's expiration date is 07/31/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative t, corrected data f) internal report (b)(4).Returned strips evaluated with no defect found.Meter not returned for evaluation.Wrong meter returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from results obtained of 47 and 53mg/dl.The customer's expected fasting blood glucose test result range is 120 - 160mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer was not able to perform a back to back blood test during the call on (b)(6) 2018,.The product is stored according to specification in the dining room.The test strip lot manufacturer's expiration date is 07/31/2019 and open vial date is july 2018.The meter memory was reviewed for previous test result history: result 1 :47mg/dl date:(b)(6) 2018, time:7:39am fasting, result 2 :47mg/dl date:(b)(6) 2018, time:7:39am fasting, result 3 :117mg/dl date:(b)(6) 2018, time:9:59am fasting, result 4 :120mg/dl date:(b)(6) 2018, time:9:54am fasting, result 5 :53mg/dl date:(b)(6) 2018, time:9:54am fasting.
|
| |
|