| Model Number |
84192961 |
| Medical Device Problem Codes |
Product Quality Problem (1506); Defective Device (2588); Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
08/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Bayer product analysis received and examined two returned syringes from the syringe kit.Visual examination confirmed the presence of a particle within the syringe barrel of one syringe.The second syringe showed no evidence of any particulate present.The particle was found to be a translucent, plastic-like material.Our investigation determined that the particle noted in the syringe was introduced during the manufacturing process.A 12 month review of complaint history determined the complaint rate to be (b)(4).
|
| |
|
Event or Problem Description
|
|
The site reported the following: white particles were seen in the syringe barrel of a sds-ctp-spk syringe kit.The kit was removed from the area and was not used for a patient.
|
| |
|