| Model Number |
8DIC |
| Medical Device Problem Code |
Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
Extubate (2402)
|
| Date of Event |
08/13/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the unit's inner cannula had a piece of round plastic in it that was blocking it.The plastic is connected in the tube and won't come out without force.The patient was recannulated.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample was received for evaluation.A visual inspection was performed, and it was observed with a piece of plastic, the failure mode occlusion was confirmed.If information is provided in the future, a supplemental report will be issued.
|
| |
|