| Catalog Number |
515301 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device expiration date: unknown.Device manufacture date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd phaseal¿ secondary set drip chamber c60 are leaking where the bag is spiked.There was no report of exposure, serious injury or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: as neither a lot number nor a sample were available for this incident, a complete investigation consisting of a sample analysis and a production history review could not be performed.The connector component is manufactured within a bd facility.During the connector molding process and assembly process, regular inspections are performed to avoid faulty product.The secondary set is manufactured at a subcontractor location.After manufacturing is completed, final samples are sent to a bd facility for quality testing.A complaint has been sent to our subcontractor location regarding this incident.Based on the limited investigation, a definite cause for the reported incident could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that bd phaseal¿ secondary set drip chamber c60 are leaking where the bag is spiked.There was no report of exposure, serious injury or medical intervention.
|
| |
|