PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 29901A
Device Problems Computer Software Problem (1112); Computer Operating System Problem (2898); Failure to Shut Off (2939)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/13/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis was unable to replicate the reported experience.However, analysis noted errors in the device log.The pr ogrammer software was reloaded.The device passed final functional and system tests.No other anomalies were found.E3: non-healthcare professional.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer displayed an error code and the screen froze.The user was unable to turn off the programmer.The battery was removed to reset the programmer.The user attempted to check a patient again after the reset and the screen froze and displayed an error code and was unable to complete the testing.The programmer was returned for service.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7828642
MDR Text Key118751355
Report Number3004593495-2018-00808
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00643169763548
UDI-Public00643169763548
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number29901A
Device Catalogue Number29901A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Date Manufacturer Received08/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-