• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM TITAMAX IMPLANT 3.75X15 MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM TITAMAX IMPLANT 3.75X15 MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.611
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 06/20/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.In telephone contact, the complainant informed that the information about lot number of the product was not available.Considering the surgical methodology, it was seen that the dentist has used less drills than recommended to perform the implant installation.
 
Event or Problem Description
(b)(4) ¿ the dentist reported that 7 days after dental implant was installed in patient¿s mouth, its non-osseointegration was observed.55 ncm of primary stability was achieved, the patient presented bone type i and immediate load and immediate implant procedures were performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM TITAMAX IMPLANT 3.75X15 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7828828
Report Number3008261720-2018-03215
Device Sequence Number1517702
Product Code DZE
UDI-Device Identifier07898237568638
UDI-Public07898237568638
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K101945
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/29/2018
Report Date (Section F) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number109.611
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/29/2018
Event Location Other
Date Report to Manufacturer06/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient Weight84
-
-