| Catalog Number |
109.328 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.
|
| |
|
Event or Problem Description
|
|
(b)(4).The dentist reported that 5 months after dental implant was installed in intraoral region 14#, its non-osseointegration was observed.A 30ncm of primary stability was achieved, the patient presented bone type iii and bone graft was executed.
|
| |
|