Catalog Number 109.328
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Tissue Damage (2104)
Date of Event 06/29/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has used less drills than recommended to perform the implant installation.
 
Event or Problem Description
(b)(6) ¿ the dentist reported that 24 days after dental implant was installed in intraoral region 18#, its non-osseointegration was observed.40 ncm of primary stability was achieved, immediate load was performed and the patient presented bone type ii.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI TITAMAX IMPLANT (5.0)4.5X11 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7829168
Report Number3008261720-2018-03498
Device Sequence Number2336762
Product Code DZE
UDI-Device Identifier07898237567136
UDI-Public07898237567136
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K101207
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2015
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/29/2018
Report Date (Section F) 07/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/07/2014
Device Catalogue Number109.328
Device Lot Number8017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/04/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/29/2018
Device Age56 MO
Event Location Other
Date Report to Manufacturer07/04/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient Weight86
-
-