| Catalog Number |
109.687 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Pain (1994); Tissue Damage (2104)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reports 5 months after the dental implant and abutment were placed in (b)(6) site 3 in the mouth, the implant failed in type ii bone.Clinician reports pain.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has used less drills than recommended to perform the implant installation.
|
| |
|
Event or Problem Description
|
|
The clinician reports 5 months after the dental implant and abutment were placed in ada site 3 in the mouth, the implant failed in type ii bone.Clinician reports pain.The product will be forwarded to the manufacturer for investigation.There were no reported post-operative complications.(b)(4).
|
| |
|