Catalog Number 09D28-03
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product recall letter was issued to all architect customers who have received the architect ict module, list number 09d28-03, lot number 180326.The letter informs the customer of the issue regarding higher than expected serum or plasma chloride results for quality control and patient samples.The letter instructs the customer to discontinue use of the suspect lot and destroy any remaining inventory.The cause of the upward shift has been traced to an omitted process step during the manufacture of the chloride electrode element of the ict module.
 
Event or Problem Description
Abbott laboratories has confirmed higher than expected serum or plasma chloride results when using architect ict module, ln 09d28-03, lot number 180326.Based on an analysis of qc data from existing customers, the results generated using this lot number may be up to 6.2% higher than expected.This qc analysis does not indicate an issue with urine chloride results.Additionally, since this issue only impacts the chloride electrode element of the ict module, potassium and sodium results are not impacted.There has been no report of impact to patient management due to this issue.However, there is a potential to generate falsely elevated serum or plasma chloride results.There is also the potential to delay the reporting of chloride ict results due to the time required for the user to attempt troubleshooting of the issue with falsely elevated serum chloride quality control results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT ICT MODULE
Common Device Name
ICT MODULE
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
MDR Report Key7830570
Report Number1628664-2018-01620
Device Sequence Number16755884
Product Code CGZ
UDI-Device Identifier00380740019327
UDI-Public00380740019327
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K980367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/26/2018
Device Catalogue Number09D28-03
Device Lot Number180326
Was Device Available for Evaluation? Yes
Type of Report(Section G)Five-Day
Initial Date Received by Manufacturer 08/24/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1628664-08/29/18-003-R
Patient Sequence Number1
-
-