• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY SEE H-10; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY SEE H-10; INTRAOCULAR LENS Back to Search Results
Model Number CC4204A
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical product: collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.The lens was returned in liquid, in lens case/vial.Visual inspection found the lens torn into 2 pieces, from one haptic to the other haptic.The injector was returned and the plunger was found to be bent.One similar complaint was reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated a cc4204a collamer single piece lens, +23.5 diopter, was torn during the loading process and there was no patient contact.The backup lens was implanted with no problem.The reporter indicated the cause of the event was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEE H-10
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7830749
Report Number2023826-2018-01297
Device Sequence Number1516977
Product Code HQL
UDI-Device Identifier00841542102469
UDI-Public00841542102469
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P990013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Model NumberCC4204A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2018
Initial Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-