| Model Number |
CC4204A |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical product: collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.The lens was returned in liquid, in lens case/vial.Visual inspection found the lens torn into 2 pieces, from one haptic to the other haptic.The injector was returned and the plunger was found to be bent.One similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated a cc4204a collamer single piece lens, +23.5 diopter, was torn during the loading process and there was no patient contact.The backup lens was implanted with no problem.The reporter indicated the cause of the event was unknown.
|
| |
|
Search Alerts/Recalls
|