Model Number 000333
Medical Device Problem Codes Fitting Problem (2183); Defective Component (2292)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the large cap on the dual port feeding adapter became loose one month after placement and cap was easily removed.There was no reported patient injury.
 
Event or Problem Description
It was reported that the large cap on the dual port feeding adapter became loose one month after placement and cap was easily removed.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one dualport feeding adaptor was returned for evaluation.Functional, visual and dimensional evaluations were performed.The investigation is confirmed for a loose cap as the large port inner and outer diameters, as well as the small port cap outer diameter, were found to be somewhat above specification.No obvious damage was apparent from the sample evaluation photo.The root cause was determined to be supplier related.Labeling review:a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL PORT FEEDING ADAPTER 20F
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7830846
Report Number3006260740-2018-02263
Device Sequence Number1286923
Product Code KNT
UDI-Device Identifier10801741080934
UDI-Public(01)10801741080934
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K915841
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000333
Device Catalogue Number000333
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Initial Date Received by Manufacturer 08/06/2018
Supplement Date Received by Manufacturer05/20/2019
Initial Report FDA Received Date08/29/2018
Supplement Report FDA Received Date05/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-