| Catalog Number |
3549216 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A device history record review is in progress.Once completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) female patient underwent a primary breast reconstruction procedure with a mentor cpx 4 breast tissue expander with suture tabs 800cc device that deflated after implantation.As a result, the patient underwent explantation and replacement with a mentor cpx 4 breast tissue expander with suture tabs 800cc device on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
On 08/30/2018, the device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.On 09/11/2018, mentor completed the investigation on the suspect medical device.Upon receipt by mentor, the device returned without fluid.No foreign material was observed within the device or on the shell surface.The product evaluation (pe) team discovered a rent at the patch junction.Microscopic examination of the edges of the rent gave no indications as to cause.No other anomalies were discovered.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet; however, deflation trends for mentor saline-filled devices will continue to be monitored by quality assurance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|