| Model Number |
MMT-723NAL |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump was unable to preparing to prime.Customer blood glucose level was160mg/dl.Troubleshooting was not done.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with broken reservoir tube lip and missing reservoir tube o-ring.Unable to prime during prime test due to faulty force sensor resistor (gold).Unable to perform excessive no delivery test due to priming anomaly.Unable to perform displacement test, basic occlusion test and occlusion test due to broken reservoir tube lip.A test reservoir was installed and did not lock in place due to broken reservoir tube lip.
|
| |
|