Model Number MMT-723NAL
Medical Device Problem Code Priming Problem (4040)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump was unable to preparing to prime.Customer blood glucose level was160mg/dl.Troubleshooting was not done.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with broken reservoir tube lip and missing reservoir tube o-ring.Unable to prime during prime test due to faulty force sensor resistor (gold).Unable to perform excessive no delivery test due to priming anomaly.Unable to perform displacement test, basic occlusion test and occlusion test due to broken reservoir tube lip.A test reservoir was installed and did not lock in place due to broken reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7831441
Report Number2032227-2018-20780
Device Sequence Number1263605
Product Code OYC
UDI-Device Identifier00613994904331
UDI-Public(01)00613994904331
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAL
Device Catalogue NumberMMT-723NAL
Device Lot NumberA5723NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2019
Initial Date Received by Manufacturer 08/07/2018
Supplement Date Received by Manufacturer04/10/2019
Initial Report FDA Received Date08/29/2018
Supplement Report FDA Received Date04/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age29 YR
Patient Weight290
-
-