C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP¿ FOLEY TRAY SYSTEM BARDEX® I.C. COMPLETE CARE FOLEY CATHETER
|
Back to Search Results |
|
| Model Number |
A319414AM |
| Medical Device Problem Codes |
Gel Leak (1267); Misassembled (1398)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that when the foley tray was opened, the saline syringe was found to be missing the saline.Allegedly, there was no cap on the end of the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed.Visual evaluation of the returned sample noted one opened, unused foley tray product present.Visual inspection of the returned tray noted the water syringe in the tray was out of place and completely empty upon sample receipt.No cap was present on the end of the syringe or within the tray, which was considered out of specification (pre-filled water syringe must not leak and must have cap in place).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use was found adequate and states the following: "proper techniques for urinary catheter insertion: perform hand hygiene immediately before and after insertion, insert urinary catheters using aseptic technique and sterile equipment, use the smallest foley catheter possible, consistent with good drainage, document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record.Proper techniques for urinary catheter maintenance: secure the foley catheter, use the statlock® foley stabilization device if provided.Maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.Maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.Keep the collection bag below the level of the bladder or hips at all times empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate leave foley catheter in place only as long as needed.1.Wash hands and don clean gloves.2.Explain procedure to patient and open peri-care kit.3.Use the provided packet of wipes to cleanse patient¿s periurethral area.4.Remove gloves and perform hand hygiene with provided alcohol hand sanitizer gel 5.Using proper aseptic technique open csr wrap.6.Don sterile gloves.7.Place underpad beneath patient, plastic/¿shiny¿ side down note: use caution to maintain aseptic technique.8.Position fenestrated drape on patient.9.Saturate 3 foam swab sticks in povidone iodine.10.Attach the water filled syringe to the inflation port note: it is not necessary to pre-test the foley catheter balloon.11.Remove foley catheter from wrap and lubricate catheter.12.Prepare patient with 3 foam swab sticks saturated in povidone iodine.Use the nondominant hand for the genitalia and the dominant hand for the swabs.Note: use each swab stick for one swipe only.Female patient: with a downward stroke cleanse the right labia minora and discard the swab.Do the same for the left labia minora.With the last swabstick cleanse the middle area between the labia minora.Male patient: cleanse the penis in a circular motion starting at the urethral meatus and working outward.13.Proceed with catheterization in usual manner using the dominant hand.A.When catheter tip has entered bladder, urine will be visible in the drainage tube.B.Insert catheter two more inches and inflate catheter balloon.14.Inflate catheter balloon using entire 10cc of sterile water provided in the prefilled syringe.Note: use of less than 10cc can result in asymmetrically inflated balloon.15.Once inflated, gently pull catheter until the inflated balloon is snug against the bladder neck.16.Secure the foley catheter to the patient use the statlock® foley stabilization device.If provided (see statlock® foley stabilization device ifu).Note: please make sure patient is appropriate for use of statlock® stabilization device.17.Position hanger on bed rail at the foot of the bed.Note: exercise care to keep bag off the floor.18.Use green sheeting clip to secure drainage tube to the sheet.Make sure tube is not kinked.19.Indicate time and date of catheter insertion on provided labels.Place designated labels on patient chart and drainage system.20.Document procedure according to hospital protocol foley catheter removal.1.To deflate catheter balloon: gently insert a luer lock or slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.2.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.3.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.4.If the balloon will not deflate and if permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.5.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient¿.
|
| |
|
Event or Problem Description
|
|
It was reported that when the foley tray was opened, the saline syringe was found to be missing the saline.Allegedly, there was no cap on the end of the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed.Visual evaluation of the returned sample noted one opened, unused foley tray product present.Visual inspection of the returned tray noted the water syringe in the tray was out of place and completely empty upon sample receipt.No cap was present on the end of the syringe or within the tray, which was considered out of specification (pre-filled water syringe must not leak and must have cap in place).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use was found adequate and states the following: "proper techniques for urinary catheter insertion ¿ perform hand hygiene immediately before and after insertion ¿ insert urinary catheters using aseptic technique and sterile equipment ¿ use the smallest foley catheter possible, consistent with good drainage ¿ document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record proper techniques for urinary catheter maintenance ¿ secure the foley catheter, use the statlock® foley stabilization device if provided ¿ maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions ¿ maintain unobstructed urine flow and keep the catheter and collection tube free from kinking ¿ keep the collection bag below the level of the bladder or hips at all times ¿ empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient ¿ routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate ¿ leave foley catheter in place only as long as needed 1.Wash hands and don clean gloves 2.Explain procedure to patient and open peri-care kit 3.Use the provided packet of wipes to cleanse patient¿s periurethral area 4.Remove gloves and perform hand hygiene with provided alcohol hand sanitizer gel 5.Using proper aseptic technique open csr wrap 6.Don sterile gloves 7.Place underpad beneath patient, plastic/¿shiny¿ side down note: use caution to maintain aseptic technique 8.Position fenestrated drape on patient 9.Saturate 3 foam swab sticks in povidone iodine 10.Attach the water filled syringe to the inflation port note: it is not necessary to pre-test the foley catheter balloon 11.Remove foley catheter from wrap and lubricate catheter 12.Prepare patient with 3 foam swab sticks saturated in povidone iodine.Use the nondominant hand for the genitalia and the dominant hand for the swabs note: use each swab stick for one swipe only female patient: with a downward stroke cleanse the right labia minora and discard the swab.Do the same for the left labia minora.With the last swabstick cleanse the middle area between the labia minora male patient: cleanse the penis in a circular motion starting at the urethral meatus and working outward 13.Proceed with catheterization in usual manner using the dominant hand a.When catheter tip has entered bladder, urine will be visible in the drainage tube b.Insert catheter two more inches and inflate catheter balloon 14.Inflate catheter balloon using entire 10cc of sterile water provided in the prefilled syringe note: use of less than 10cc can result in asymmetrically inflated balloon 15.Once inflated, gently pull catheter until the inflated balloon is snug against the bladder neck 16.Secure the foley catheter to the patient use the statlock® foley stabilization device if provided (see statlock® foley stabilization device ifu) note: please make sure patient is appropriate for use of statlock® stabilization device 17.Position hanger on bed rail at the foot of the bed note: exercise care to keep bag off the floor 18.Use green sheeting clip to secure drainage tube to the sheet.Make sure tube is not kinked 19.Indicate time and date of catheter insertion on provided labels.Place designated labels on patient chart and drainage system 20.Document procedure according to hospital protocol foley catheter removal 1.To deflate catheter balloon: gently insert a luer lock or slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve 2.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently 3.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation 4.If the balloon will not deflate and if permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol 5.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient¿.
|
| |
|
Event or Problem Description
|
|
It was reported that when the foley tray was opened, the saline syringe was found to be missing the saline.Allegedly, there was no cap on the end of the syringe.
|
| |
|