MAUDE Adverse Event Report: WARSAW ORTHOPEDICS ATLAS CABLE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS ATLAS CABLE SYSTEM Back to Search Results
Catalog Number 826-213
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-op diagnosis :humerus patellar fracture procedure: fracture it was reported that the cable broke intra-operatively.Patient was involved in the event but no patient complications or symptoms occurred.
 
Additional Manufacturer Narrative
Product analysis: the cable has been returned in two sections.A visual review of the two severed ends indicates the two pieces do not match.As one end appears to be a clean cut and the other end appears to have been frayed.A microscopic review of the individual cables shows that the angles are similar and there does not appear to be necking that is associated with linear force.There is also damage to the eyelet noted.It appears the cable may have been cut.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS CABLE SYSTEM
Common Device Name
CABLE
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7831744
Report Number1030489-2018-01180
Device Sequence Number1271711
Product Code ISN
UDI-Device Identifier00673978047980
UDI-Public00673978047980
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K892651
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2025
Device Catalogue Number826-213
Device Lot Number0603584W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer10/19/2018
Initial Report FDA Received Date08/30/2018
Supplement Report FDA Received Date10/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
-
-