| Model Number |
MMT-551NAB |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they received no delivery alarm after changing set and reservoir blood glucose was 476 mg/dl e of incident.The customer troubleshooted and was advised to set change and reservoir change.The insulin pump will need to be replaced.The customer was advised to revert to back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No excessive no delivery alarm and no device deficiency anomaly noted during test.(b)(4).
|
| |
|