| Catalog Number |
222343 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/26/2014
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the (b)(4) complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported by the hospital that the surgeon was doing a remplissage procedure for a shoulder repair using the healix br 3.4 mm anchor and passing suture retrograde with the ideal suture graspers.The first healix br 3.4mm anchor he went to implant and the tip of the anchor snapped just trying to push the anchor through the tissue to get to the bone hole.It was reported that it was suggested to open and use the percannula system to make this procedure and implanting the anchors easier and the surgeon agreed so we opened the percannula and implanted a healix br 3.4mm anchor with no issues.We then opened a 30 degree ideal suture grasper and when we had the suture hooked in the retrieval clip and went to pull it through the tissue the clip did not hang on to the stitch and straightened out making the tool unusable.It was reported that a 15 degree ideal suture grasper was opened to try as the surgeon thought that angle might be better anyway and the same thing happened.It was reported that the surgeon realized the clips on these are fragile, but for that to happen on one pass each was not acceptable to him.We opened one more 15 degree ideal suture grasper, i got him to slow down and listen to my advice and once he had the suture in the grasper to put it in ¿slide¿ mode and gently toggle the tip of the instrument back through the tissue.This technique worked great and we passed the sutures from the healix 3.4 anchor with ease from there on.We then located the stitches in the subacromial space, tied them and completed the repair very nicely.It was not reported if there were any delays in the surgical procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|